Predicted Ranges and Resolution
The graphs below demonstrate the range and resolution of Ocean Optics spectrometers.

USB Series

USB Bench with Sony Detector: USB2000+
  Example: If the starting wavelength is 350 nm, then the range is ~674 nm, providing a 350-1024 nm wavelength range and 1.38 nm resolution.
  Example: If the starting wavelength is 400 nm, then the range is ~316 nm, providing a 400-716 nm wavelength range and 0.65 nm resolution.
  Example: If the starting wavelength is 400 nm, then the range is ~195 nm, providing a 400-595 nm wavelength range and 0.40 nm resolution.
  Example: If the starting wavelength is 350 nm, then the range is ~137 nm, providing a 350-487 nm wavelength range and 0.28 nm resolution.
USB Bench with Toshiba Detector: USB4000
  Example: If the starting wavelength is 350 nm, then the range is ~692 nm, providing a 350-1042 nm wavelength range and 1.40 nm resolution.
  Example: If the starting wavelength is 400 nm, then the range is ~325 nm, providing a 400-725 nm wavelength range and 0.67 nm resolution.
  Example: If the starting wavelength is 400 nm, then the range is ~195 nm, providing a 400-595 nm wavelength range and 0.40 nm resolution.
  Example: If the starting wavelength is 350 nm, then the range is ~140 nm, providing a 350-490 nm wavelength range and 0.29 nm resolution.

HR Series

HR Bench with Sony Detector: HR2000+
  Example: If the starting wavelength is 400 nm, then the range is ~437 nm, providing a 400-837 nm wavelength range and 0.32 nm resolution.
  Example: If the starting wavelength is 600 nm, then the range is ~188 nm, providing a 600-788 nm wavelength range and 0.137 nm resolution.
  Example: If the starting wavelength is 700 nm, then the range is ~90 nm, providing a 700-790 nm wavelength range and 0.065 nm resolution.
  Example: If the starting wavelength is 400 nm, then the range is ~82 nm, providing a 400-482 nm wavelength range and 0.06 nm resolution.
HR Bench with Toshiba Detector: HR4000
  Example: If the starting wavelength is 400 nm, then the range is ~437 nm, providing a 400-837 nm wavelength range and 0.239 nm resolution.
  Example: If the starting wavelength is 300 nm, then the range is ~215 nm, providing a 300-515 nm wavelength range and 0.117 nm resolution.
  Example: If the starting wavelength is 700 nm, then the range is ~90 nm, providing a 700-790 nm wavelength range and 0.05 nm resolution.
  Example: If the starting wavelength is 400 nm, then the range is ~82 nm, providing a 600-682 nm wavelength range and 0.045 nm resolution.

QE Series

  Example: If the starting wavelength is 400 nm, then the range is ~364 nm, providing a 400-764 nm wavelength range and 0.645 nm resolution.
  Example: If the starting wavelength is 600 nm, then the range is ~152 nm, providing a 600-752 nm wavelength range and 0.295 nm resolution.
  Example: If the starting wavelength is 300 nm, then the range is ~112 nm, providing a 300-412 nm wavelength range and 0.226 nm resolution.
  Example: If the starting wavelength is 450 nm, then the range is ~62 nm, providing a 450-512 nm wavelength range and 0.139 nm resolution.

NIRQuest Series

  Example: If the starting wavelength is 1000 nm, then the range is ~1662 nm, providing a 1000-2662 nm wavelength range and 5.34 nm resolution.
  Example: If the starting wavelength is 1300 nm, then the range is ~1236 nm, providing a 1300-2536 nm wavelength range and 3.98 nm resolution.
  Example: If the starting wavelength is 1300 nm, then the range is ~815 nm, providing a 1300-2115 nm wavelength range and 2.63 nm resolution.
  Example: If the starting wavelength is 1700 nm, then the range is ~380 nm, providing a 1700-2080 nm wavelength range and 1.23 nm resolution.
  Example: If the starting wavelength is 1700 nm, then the range is ~270 nm, providing a 1700-1970 nm wavelength range and 0.85 nm resolution.
  Example: If the starting wavelength is 1900 nm, then the range is ~192 nm, providing a 1900-2092 nm wavelength range and 0.62 nm resolution.
  Example: If the starting wavelength is 1700 nm, then the range is ~154 nm, providing a 1700-1854 nm wavelength range and 0.49 nm resolution.
  Example: If the starting wavelength is 1300 nm, then the range is ~74 nm, providing a 1300-1374 nm wavelength range and 0.24 nm resolution.